Models TRID30-7R : (replaceable battery, download instructions here)
TRID30-7F : (fixed battery)
TRID30-7FW (fixed battery, preconfigured and package to meet WHO specification E006/TR06.3)
Measurement Range -30°C ~ +60°C (-22°F ~ +140°F)
Resolution 0.1°C(0.2°F) for measurements -30°C ~ +40°C (-22°F ~ +104°F)
0.2°C(0.4°F) for measurements above +40°C (+104°F)
Accuracy Better than ±0.5°C (±0.9°F) for measurements from -20°C~+40°C (-4°F~+104°F) -typically ±0.3°C (0.6°F)
Better than ±0.8°C (±1.5°F) for other measurements – typically ±0.5°C (0.9°F).
Sensor Reaction Time Typically less than 5 minutes (T90) in moving air (1m/s).
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values
Sampling Frequency Adjustable, 30 seconds to several hours
Logging Start Options Push button start or specific date & time. Optional start delay of up to 18hours.